IN VITRO EVALUATION OF THE EFFECTIVENESS OF THE APICAL LOCATOR ROOT ZX IN PRESENCE OF SODIUM HYPOCHLORITE COMBINED WITH RC - PREP OR EDTA COMPARED WITH SCANNING ELECTRON MICROSCOPE

Authors

  • Monique Marie Gay Ortiz U. Santo Tomás
  • Martha Elena Varón Plata U. Santo Tomás
  • Oscar Gustavo Serrano Navas U. Santo Tomás

DOI:

https://doi.org/10.15332/us.v2i2.1845

Keywords:

Root ZX, Cement-Dentin-Canal junction (CDC), sodium hypochlorite, Rc - Prep, EDTA

Abstract

Purpose: To assess the effectiveness of the Root Zx to determine the Cement - Dentin - Canal junction (CDC) in the presence of sodium hypochlorite 5,25% Clorox®; combined with Rc- prep. Premier® and sodium hypochlorite 5.25% Clorox® combined with EDTA 10% Eufar® in uniradicular teeth.
Material and methods: An experimental study in vitro with 12 uniradicular teeth was done. They wered divided in three groups: Grupo 1 (n = 12) sodium hypochlorite 5,25% Clorox ®; Group 2 (n = 6) sodium hypochlorite 5.25%Clorox® combined with Rc- Prep. Premier®; Group 3 (n = 6) sodium hypochlorite 5,25% Clorox ® combined with EDTA 10% Eufar®. Root ZX was used to locate the CDC; the samples were divided longitudinally, covered with gold and observed at the SEM, to corroborate the effectiveness of the instrument in the location of the CDC.
Results: The results demonstrate statistically significant differences between Root ZX and SEM in each group. In group 1, a longer measurement was obtained with the Root ZX compared with the SEM, the difference was 0.788 mm. On the other hand, in the group of the sodium hypochlorite 5,25% with EDTA 10%, a shorter measurement was obtained, (-0.766 mm) a statistically significant difference.
Conclusions: The results suggest that the Root ZX is effective in the location of the apical constriction with a level of tolerance of +/- 1 mm.

Downloads

Download data is not yet available.

Author Biographies

Monique Marie Gay Ortiz, U. Santo Tomás

Odontóloga, U. Javeriana, Residente II año, posgrado de Endodoncia, U. Santo Tomás

Martha Elena Varón Plata, U. Santo Tomás

Odontóloga, Residentes II año posgrado de Endodoncia, U. Santo Tomás

Oscar Gustavo Serrano Navas, U. Santo Tomás

Odontólogo, Residentes II año posgrado de Endodoncia, U. Santo Tomás

References

West J, Roane J. Sistema de limpieza y conformación de los canales radiculares. En: Cohen S, Burns R. Vías de la Pulpa. 7ma ed. Madrid: Editorial Harcourt- Mosby; 1999. p. 209 - 211.

Nahmias J, Aurelio J, Gerstein H. An in vitro model for evaluation for electronic root canal length measuring devices. J Endod 1987; 13: 209 - 214.

Walton Torabinejad M. Endodoncia principios y Práctica. 2nda ed. México: Mc Graw-Hill; 1996.

McDonald NJ, Hovland EJ. An evaluation of the apex locator Endocater. J Endod 1990; 16: 5 - 8.

Kuttler Y. Microscopic investigation of root apexes. J Am Dent Assoc.1955; 50: 544 - 552.

Hembrough JH, Weine FS, Pisano JV, Eskoz N. Accuracy of an electronic apex locator: A clinical evaluation in maxillary molars. J Endod 1993; 19: 242 - 246.

Ibarrola JL, Chapman B, Howard JH, Knowles KI. Effect of preparing on root ZX apex locators. J Endod 1999; 25: 625 - 626.

Jenkins JA, Walker WA. An in vitro evaluation of the accuracy of the Root ZX in presence of various irrigants. J Endod 2001; 27: 209 - 211.

Imao S. Nuevo método para medir la longitud de trabajo. J Dent Res 1962; 41: 375 - 387.

Shabahang S, Goon WY, Gluxkin AH. An in vivo evaluation of Root ZX electronic apex locator. J Endod 1996; 22: 616 - 617.

Pratten DH, McDonald NJ. Comparison of radiographic and electronic working lengths. J Endod 1996; 22: 173 -176.

Kobayashi C. New electronic canal measuring device based on the ratio method. J Endod 1994; 20: 111 - 114.

Frank AL, Torabinejad M. An in vivo evaluation of endex apex locator. J Endod 1993; 19: 177 - 179.

Bramamnte C. A critical evaluation of some methods of determining tooth length. Oral Surg Oral Med Oral Pathol 1974; 37: 463 - 469.

Weiger R, John X, Geigle H y Löst C. An in vitro comparison of two modern apex locators. J Endod 1999; 25: 765 - 768.

Donnelly JC. Simplified model to demonstrate the operation of electronic root canal measuring devices. J Endod.1993; 19: 579 - 580.

Kaufman, Katz J. Reliability of root ZX apex locator tested by an in vitro model. J Endod 1993; 19: 20 [Abstract].

Ounsi HF, Naaman A. In vitro evaluation of reliability of the Root ZX electronic apex locator. Int J Endod 1999; 32: 120 - 123.

Pagano M, Gauvreau K. Principles of biostatistics. Belmont: Duxbury Press; 1993.

Bland E, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 8: 307 - 310.

Ponce E, Fernández JA. The cemento dentino canal - junction, the apical foramen and the apical constriction: Evaluation by optical microscopy. J Endod 2003; 29: 214 - 219.

Rivas Muñoz R. Uso y eficacia del localizador del foramen fisiológico, apex finder 7001.URL disponible en: http://www.iztacala.unam.mx/~rrivas/apexfinder.html.

Keller ME, Brown CE J, Newton CW. A clinical evaluation of Endocater - An electronic apex locator. J Endod. 1991; 17: 271 - 274.

Fouad AF, Rivera EM, Krell KV. Accuracy of endex with variations in canal irrigants and foramen size. J Endod 1993; 19: 63 - 67.

Mayeda DL, Simon JH, Aimar DF, Finley K. In vivo measurements accuracy in vital and necrotic canals with the endex apex locator. J Endod 1993; 19: 545 - 548.

Kobayashi C. Electronic canal length measurement. Oral Surg Oral Med Oral Pathol 1995; 79: 226 -231.

Ounsi HF, Haddad G. In vitro evaluation of the reliability of the endex electronic apex locator. J Endod 1998; 24:120 - 121.

Pagavino G, Pace R, Baccetti T. A SEM study of in vivo accuracy of the Root ZX electronic apex locator. J Endod 1998; 24: 438 - 441.

Cruz-Filho AM, Sousa-Neto MD, Saquy PC, Pecora JD. Evaluación del efecto del EDTAC, CDTA Y EGTA sobre la microdureza dentinal. J Endod 2001; 27: 183 - 184.

Pilot TF, Pitts DL. Determination of impedance changes at varying frequencies in relation to root canal position and irrigant. J Endod 1997; 23: 719 - 724.

Tinaz AC, Sevimli LS, Gorgul G, Turkoz EG. The effects of sodium hypochlorite concentrations on the accuracy of an apex locating device. J Endod 2002; 28:160 - 162.

Comunicación personal con la Ingeniera Zarith Pachón, Ingeniera Química, Instituto Colombiano de Petróleos (ICP). Piedecuesta, Santander, 11 de mayo de 2003.

How to Cite

Gay Ortiz, M. M., Varón Plata, M. E., & Serrano Navas, O. G. (2018). IN VITRO EVALUATION OF THE EFFECTIVENESS OF THE APICAL LOCATOR ROOT ZX IN PRESENCE OF SODIUM HYPOCHLORITE COMBINED WITH RC - PREP OR EDTA COMPARED WITH SCANNING ELECTRON MICROSCOPE. Ustasalud, 2(2), 97–106. https://doi.org/10.15332/us.v2i2.1845

Issue

Section

Scientific and technological research papers